Test and analysis of electromagnetic compatibility of heart sound electrocardiograph detector and research on rectification method
Author:
Affiliation:

1.State Key Laboratory of Dynamic Measurement Technology,North University of China,Taiyuan Shanxi 030051,China;2.Chinese General Hospital of the People's Liberation Army,Beijing 100853,China

Funding:

Ethical statement:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
    Abstract:

    Modern medical devices are more inclined to high precision, high sensitivity, and miniaturization, resulting in more and more complex electromagnetic environments. External electromagnetic interference will also affect medical devices, and even affect doctors' diagnosis in severe cases. In this paper, combined with the heart sound electrocardiograph detector, the conduction emission test, radiation emission test and electrostatic discharge test are studied based on the medical device test standard. Some rectification methods are proposed, such as rectifying the shell, adding the filter circuit, adding the magnetic ring and adding the charge release circuit. It can be used as a reference for the analysis and rectification of other types of related medical devices.

    Reference
    Related
    Cited by
Get Citation

王朔彤,史鹏程,张国军,崔建功,力乙瑞,王博,汪涛.心音心电检测仪电磁兼容问题的测试分析及整改方法[J]. Journal of Terahertz Science and Electronic Information Technology ,2024,22(5):523~528

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
History
  • Received:September 02,2022
  • Revised:November 28,2022
  • Adopted:
  • Online: June 03,2024
  • Published: